Anti-Romiplostim (Nplate) ELISA Kit (KAD76501)

价格:
规格:
  • 96T
  • 大包装询价
数量:
品牌:

AntibodySystem

概述
  • 货号

    KAD76501

  • 描述

    This assay employs the quantitative sandwich enzyme immunoassay technique. Romiplostim has been pre-coated onto a microplate. Samples or standards are pipetted into microwells and anti-Romiplostim will be captured by immobilized Romiplostim. After washing away any unbound substances, a biotin-labeled Romiplostim is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Anti-Romiplostim bound in the initial step. The color development is stopped and the intensity of the color is measured.

  • 应用

    Used for the quantitative determination of Anti-Romiplostim concentration in serum and plasma.
  • 检测方法

    Colorimetric
  • 样本类型

    Plasma, Serum
  • 实验类型

    Quantitative
  • 检测范围

    31.25 - 2,000 ng/mL
  • 灵敏度

    17.08 ng/mL
  • 精准度

    Intra-Assay Precision (Precision within an assay): <10%

    Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision.

    Inter-Assay Precision (Precision between assays): <15%

    Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.

     

    Intra-Assay Precision

    Inter-Assay Precision

    Sample

    1

    2

    3

    1

    2

    3

    n

    16

    16

    16

    24

    24

    24

    Mean (ng/mL)

    848.7

    206.8

    46.8

    888.7

    225.1

    53.9

    Standard deviation

    24.6

    9.6

    4.3

    34.7

    13.0

    2.8

    CV (%)

    2.9

    4.7

    9.1

    3.9

    5.8

    5.2

  • 回收率

    80-120%
  • 运输

    2-8 ℃
  • 稳定性和存储

    When the kit was stored at the recommended temperature for 6 months, the signal intensity decreased by less than 20%. For unopened kits, if you want to prolong the storage time, please store the Standard, Detection A, Detection B and Microplate at - 20℃, the rest reagents should be store at 4℃.
图片
参考文献
评价