货号
KVV13401
-
描述
PRINCIPLE OF THE ASSAY This assay employs the quantitative sandwich enzyme immunoassay technique. An antibody specific for Monkeypox Virus A29L Antigen has been pre-coated onto a microplate. Standards or samples are pipetted into the wells and any Monkeypox Virus A29L Antigen present is bound by the immobilized antibody. After washing away any unbound substances, a biotin-labeled Monkeypox Virus A29L Antigen antibody is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Monkeypox Virus A29L Antigen bound in the initial step. The color development is stopped and the intensity of the color is measured.
-
靶标
A29L -
检测方法
Colorimetric -
样本类型
Plasma, Serum -
实验类型
Quantitative -
检测范围
125 - 8,000 pg/mL -
灵敏度
50 pg/mL -
精准度
Intra-Assay Precision (Precision within an assay): <10%
Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays): <15%
Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Intra-Assay Precision
Inter-Assay Precision
Sample
1
2
3
1
2
3
n
16
16
16
24
24
24
Mean (pg/mL)
4617.7
881.6
217.1
5311.8
1206.5
303.9
Standard deviation
126.6
40.8
20.1
162.5
55.5
24.0
CV (%)
2.7
4.6
9.3
3.1
4.6
7.9
-
回收率
80-120% -
运输
2-8 ℃ -
稳定性和存储
When the kit was stored at the recommended temperature for 6 months, the signal intensity decreased by less than 20%. For unopened kits, if you want to prolong the storage time, please store the Standard, Detection A, Detection B and Microplate at - 20℃, the rest reagents should be store at 4℃.