货号
KAV00109
-
描述
This assay employs the quantitative indirect enzyme immunoassay technique. Recombinant SARS-CoV-2 Spike Protein (Trimer) has been pre-coated onto a microplate. Standards or samples are pipetted into the wells and any Human Anti-SARS-CoV-2 Spike Protein (Trimer) IgG present is bound by the immobilized protein. After washing away any unbound substances, a biotin-labeled Mouse Anti-Human IgG antibody is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Human Anti-SARS-CoV-2 Spike Protein (Trimer) IgG bound in the initial step. The color development is stopped and the intensity of the color is measured.
-
靶标
Spike -
应用
Used for the quantitative determination of Anti-SARS-CoV-2 Spike Protein (Trimer) Human IgG concentration in serum and plasma. -
检测方法
Colorimetric -
样本类型
Plasma, Serum -
实验类型
Quantitative -
检测范围
125 - 8,000 pg/mL -
灵敏度
54.8 pg/mL -
精准度
Intra-Assay Precision (Precision within an assay): <10%
Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision.
Inter-Assay Precision (Precision between assays): <15%
Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Intra-Assay Precision
Inter-Assay Precision
Sample
1
2
3
1
2
3
n
16
16
16
24
24
24
Mean (pg/mL)
2802.1
561.9
146.7
3509.3
606.9
161.6
Standard deviation
104.5
26.5
12.8
301.1
50.9
18.2
CV (%)
3.7
4.7
8.7
8.6
8.4
11.2
-
回收率
80-120% -
运输
2-8 ℃ -
稳定性和存储
When the kit was stored at the recommended temperature for 6 months, the signal intensity decreased by less than 20%. For unopened kits, if you want to prolong the storage time, please store the Standard, Detection A, Detection B and Microplate at - 20℃, the rest reagents should be store at 4℃.