Human Anti-Monkeypox Virus (MPXV) IgG ELISA Kit (A35R) (KAV13101)

价格:
规格:
  • 96 T
  • 大包装询价
数量:
品牌:

AntibodySystem

概述
  • 货号

    KAV13101

  • 描述

    PRINCIPLE OF THE ASSAY. This assay employs the quantitative competitive enzyme immunoassay technique. A recombinant Monkeypox virus A35R antigen has been pre-coated onto a microplate. Standards or samples are premixed with biotin-labeled antibody and then pipetted into the wells. Human Anti-Monkeypox Virus IgG (A35R) in the sample competitively binds to the pre-coated protein with biotin-labeled Human Anti-Monkeypox Virus IgG (A35R). After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in inversely proportion to the amount of Human Anti-Monkeypox Virus IgG (A35R) bound in the initial step. The color development is stopped and the intensity of the color is measured.

  • 靶标

    A35R
  • 应用

    Used for the quantitative determination of Human Anti-Monkeypox Virus IgG (A35R) concentration in serum and plasma.
  • 检测方法

    Colorimetric
  • 样本类型

    Plasma, Serum
  • 实验类型

    Quantitative
  • 检测范围

    312.5 ng/mL - 20000 ng/mL
  • 精准度

    Intra-Assay Precision (Precision within an assay): <20%

    Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision.

    Inter-Assay Precision (Precision between assays): <20%

    Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.

     

    Intra-Assay Precision

    Inter-Assay Precision

    Sample

    1

    2

    3

    1

    2

    3

    n

    16

    16

    16

    24

    24

    24

    Mean (ng/mL)

    10326.5

    3001.4

    736.1

    9833.3

    2814.0

    737.1

    Standard deviation

    805.2

    419.3

    138.3

    797.2

    282.6

    117.7

    CV (%)

    7.8

    14.0

    18.8

    8.1

    10.0

    16.0

  • 回收率

    80-120%
  • 运输

    2-8 ℃
  • 稳定性和存储

    When the kit was stored at the recommended temperature for 6 months, the signal intensity decreased by less than 20%. For unopened kits, if you want to prolong the storage time, please store the Standard, Detection A, Detection B and Microplate at - 20℃, the rest reagents should be store at 4℃.